9.2.4 IN CNF (Input Configuration)

Input configuration is used to select:
• 50 or 60 Hz line frequency [INP.1]
• slow or fast read rate [INP.2]
• unipolar or bipolar inputs
• cold junction compensation [INP.5]
INP.1 and INP.2 are related to each other. If your power requirements require
50 Hz, you can have optimum integration in FAST read mode (12/sec). In the
FAST mode, you need a jumper in the S1A position on the vertical Signal
Input Board. If you set the SLOW read rate, this jumper should be removed
to avoid overloading the integrator. SLOW read rate produces less noise.
MENU
BUTTON
MAIN MENU
"IN CNF":
Meter Function Menus
MIN/MAX
/MENU *
BUTTON
SUB MENU
"INP.1=0"
"INP.1=1"
"INP.2=0"
"INP.2=1"
"INP.3=0"
"INP.3=1"
"INP.4=0"
"INP.4=1"
"INP.4=2"
"INP.4=3"
"INP.5=0"
"INP.5=1"
"INP.6=0"
"INP.6=1"
"INP.7=0"
"INP.7=1"
DESCRIPTION
INPUT CONFIGURATIONS
Line Frequency:
60 Hz
50 Hz
Read Rate:
Slow
Fast
Input Voltage:
Unipolar
Bipolar
Transmitter Type:
No Transmitter
Not used
Not used
Not used
Bridge Mode:
Normal Operation
Setpoint 1 value would be lower
overload limit and Setpoint 2 value
would be upper overload limit
Disabled
Enabled
Type of Reading:
Non Ratiometric
Ratiometric
9
9-11