18.1 CURRENT INPUT

INPUT RANGES(+10%):
RESOLUTION:
MAXIMUM INPUT:
INPUT OHMS:
SENSOR EXCITATION:

18.2 VOLTAGE INPUT

INPUT RANGES:
RESOLUTION:
MAX INPUT:
INPUT OHMS:
BIAS AMPS:
SENSOR EXCITATION:

18.3 POTENTIOMETER INPUT

INPUT RANGES:
RESOLUTION:
SENSOR EXCITATION:

18.4 GENERAL

SCALE:
OFFSET:
POLARITY:
NOISE REJECTION
ACCURACY at 25 ºC:
MAX ERROR:
SPAN TEMPCO:
STEP RESPONSE:
WARMUP:
UNIPOLAR:
BIPOLAR:
NMR:
CMR:
CMV:
0-20 mA and 4-20 mA
1µA
200 mA
5
10V, to 30 mA
24V, to 25 mA
100 mV
1 V
±50 mV
±0.5 V
1 µV
10 µV
70 Vp
350 Vp
1G
1M
50 pA
5 pA
10V, to 30 mA
24V, to 25 mA
1 V or 10 V
programmed to 0.001%
1.5 to 11 Vdc, to 60 mA max.
24V, to 25 mA max.
+0.000001 to +500000 or
–0.0001 to –99999.
Zero to +999999
Automatic
60 dB, 50 or 60 Hz, + selected filter
120 dB
Dielectric strength to 1500 V transient per
260 VRMS or DC working voltage.
±0.005% of reading
less than 20 ppm/ºC
1 second to 99.9%
55 minutes to rated accuracy. It is
recommended that the unit be continuously
running to insure its accuracy.
Specifications
10 V
100 V
±5 V
±50 V
100 µV
1 mV
350 Vp
300 Vp
1M
1M
1 pA
1 pA
18
18-1