Omega DP461-S User Manual page 54

High performance strain gage indicator
Table of Contents

Advertisement

8
METHOD III: SCALING BY SIMULATING A KNOWN LOAD WITH THREE OR MORE
Scaling to
Display
Engineering
Units
MENU
SUBMENU 1 SUBMENU 2 DESCRIPTION
IN.SC.OF INPUT 1
READ 1
INPUT2
READ 2
DEC PT
FFFF.FF
CNT BY
The unit is now completely scaled. If during verifying the calibra-
tion, the display is not correct, you can use the offset function
("RDG OF" if all values read high or low) or perform this calibra-
tion again.
DECIMAL POINT POSITIONS.
EXAMPLE: An end user has a 5 pound load cell with a
2 mV/V output. The end user wants to display in ten thousandths
of a pound (desired display 0.0000 to 5.0000). Proceed to Section
8.2 performing A, B and C sections.
1. Prior to simulating a known load you must set up the unit as
shown below.
Pressing the 'MENU' button will step you through all the different
menu's listed below. When you wish to enter a menu, to the
submenu 1 stage, press the 'MIN' button. To make a selection, in
submenu 1, press the 'MAX' button. If the submenu has more than
one item to select, press the 'MIN' button to scroll to the next item
of choice and again press the 'MAX' button to make your selection.
After making all your choices, in submenu 1, enter these selections
by pressing the 'MENU' button which will momentarily flash
"STORED", on the display.
If the item selected in submenu 1 has a submenu 2 associated with
it, you can press the 'MIN' button to enter the submenu 2. Select
your choice by pressing the 'MAX' button and enter your choice by
pressing the 'MENU' button.
000013
This is where you
000000
enter the data
080567
obtained during the
000160
calibration run
(see above).
Decimal point selection
Press the 'RESET' button
two times to return to
the run mode
8-10