Figure 8-6 Sensor Input Connections - Omega DP461-S User Manual

High performance strain gage indicator
Table of Contents

Advertisement

+EXCITATION
VOLTAGE

Figure 8-6 Sensor Input Connections

C. The unit comes with all lockout bytes set in the unlocked
position("0"). Refer to Section 9.1 for information on lockout
bytes.
If the unit has been programmed previously make sure that at least
the following lockout bytes are unlocked. If they are Not unlocked
you will not be able to scale the meter.
MENU
"L1CNF"
"L2CNF"
Scroll to the menu by pressing the 'MENU' button. Enter a
submenu by pressing the 'MIN' button. select "0" or "1" by pressing
the 'MAX' button. Always press the 'MENU' button when leaving
a submenu to store the values you entered.
METHOD I: SCALING WITHOUT SIMULATING A KNOWN LOAD.
NOTE: All of the following steps are accomplished by using the
front panel buttons.
1. Scale the instrument by entering only 3 parameters.
The 3 parameters are "RDG SC" (Reading Scale), "RDG OF"
(Reading Offset), and "DEC PT" (Decimal Point).
Calculate "RDG SC" using the following formula:
+OUTPUT
COMMON
JUMPER
GND
USER
PROVIDED
SUBMENU
L1C.7=0, L1C.8=0
All sub menus must be unlocked "0"
8-17
+ E
1
P9
+ S
2
– S
3
DP41
METER
– E
1
P3
NC
2
3
8
Scaling to
Display
Engineering
Units