8
Scaling to
Display
Engineering
Units
RDG SC= display span/((sensor's output span) (natural gain))
RDG SC= 4000/(4 x 10000) = 0.100
Where: display span= 4000 – 0 = 4000
Sensor's output span= 5 Vdc – 1 Vdc = 4 Vdc
Meter's natural gain= 10000 (Obtained from Table 8-1)
VOLTAGE
SENSOR'S
RANGE
OUTPUT UNITS
100 mV, ±50 mV Millivolts
1 V, ±500 mV
Millivolts
10 V, ±5 V
Volts
100 V, ±50 V
Volts
Pressing the 'MENU' button will take you through all the different
menus listed below. When you wish to enter a menu, to the
submenu 1 stage, press the 'MIN' button. To make a selection, in
submenu 1, press the 'MAX' button. If the submenu has more than
one item to select you use the 'MIN' button to scroll to the next
item of choice and again use the "'MAX'" button to make your
selection.
After making all your choices, in submenu 1, enter these selections
by pressing the 'MENU' button which will momentarily flash
"STORED" on the display. If the item selected in submenu 1 has a
submenu 2 associated with it, press the 'MIN' button to go to
submenu 2. Select your choice by pressing the 'MAX' button and
enter your choice by pressing the 'MENU' button. Pressing the
"'RESET'" button one time brings you back one 'MENU' item,
pressing it a second time returns you to the run mode.
WARNING! this also results in a hard reset of peak and valley and
your setpoints.
Proceed through the flow chart below entering the values listed in
submenu 1 and 2 columns.
8-18
METER'S
NATURAL GAIN
1000
100
10000
1000