In.sc.of (Input Scale And Offset) - Omega DP461-S User Manual

High performance strain gage indicator
Table of Contents

Advertisement

9
Meter Function Menus
* The 'MIN' button allows you to sequence through INP.1, INP.2, INP.3, INP.4,
INP.5, INP.6, and INP.7.
The 'MAX' button allows you to select the "0" or "1" state for each "INP"
condition.
The 'MENU' button stores the selected values for all "IN CNF" condition(s)
changed and advances the meter to "IN.SC.OF". Do not press the 'MENU'
button after each change within the submenu or the meter will advance to
the next menu item.
Every underlined "0" or "1" state is the factory preset value.

9.2.5 IN.SC.OF (Input Scale and Offset)

Refer to Section 8 for a detailed discussion of this feature.
Input scale and offset is typically used when you want to scale your meter
(using two input data points):
MENU
BUTTON
MAIN
MENU
"IN.SC.OF":
* The 'MIN' button allows you to sequence through "INPUT 1", "READ 1",
"INPUT 2", and "READ 2" headings.
The 'MAX' button sends you to the value that corresponds to "INPUT 1",
"READ 1", "INPUT 2", or "READ 2" so you can change it (go to the SUB
MENU 2 item).
9-12
MIN/MAX/
MIN/MAX *
MENU **
BUTTON
BUTTON
SUB MENU 1 SUB MENU 2 DESCRIPTION
INPUT 1
000000.
("000000.")
READ 1
000000.
("000000.")
INPUT 2
000000.
("100000.")
READ 2
000000.
("100000.")
INPUT SCALE OFFSET
Input scale and offset in
2-Coordinate Format
Item #1 of Coordinate #1.
Enter the first value displayed
by the meter.
Item #2 of Coordinate #1.
Enter first desired value.
Item #1 of Coordinate #2.
Enter the second
value displayed by the meter.
Item #2 of Coordinate #2.
Enter second desired value.