Omega DP461-S User Manual page 89

High performance strain gage indicator
Table of Contents

Advertisement

MENU
SUBMENU 1 SUBMENU 2 DESCRIPTION
RDG SC
0.00200
RDG OF
0000000
IN CNF
INP.1=0
INP.2=0
INP.3=0
INP.4=0
INP.5=
INP.6=0
INP.7=1
IN.SC.OF
DEC PT
FFFF.FF
CNT BY
2. Now you are in the run mode and the transducer is connected to
the meter. Simulate no load on the transducer (just leave the
pressure port open to atmosphere) and note the display reading.
Assume the display shows "1.25". To make the display indicate
zeros you must go to the menu "RDG OF" and enter the value
"–001.25". After entering this value press the 'MENU' button to
store it, and then press the 'RESET' button two times to return
to the run mode. The scaling is now complete.
METHOD II: SCALING BY SIMULATING A KNOWN LOAD.
1. Prior to simulating a known load you must set up the unit as
shown below.
The 'MENU' button will take you through all the different menu's
listed below. When you wish to enter a menu, to the submenu 1
stage, press the 'MIN' button. To make a selection, in submenu 1,
press the 'MAX' button. If the submenu has more then one item to
select use the 'MIN' button to scroll to the next item of choice and
See calculations above
60 Hz ac power
Slow reading
(S1A jumper omitted)
Unipolar input
(S1B jumper omitted)
Std, for voltage inputs
Disregard not used
Disables IN.SC.OF
Ratiometric input
Skip past
Sets decimal point
Press the 'RESET' button
two times to return to
the run mode
8-45
8
Scaling to
Display
Engineering
Units