10.2 INPUT CONFIGURATION ("INP.CNF" and "INPUt")

BIT POSITION
7
6
5
4
0
0
1
1
0
1
0
1
0
1
EXAMPLE: To set the meter with line frequency of 50 Hz, slow reading rate (in case of
MENU 2), degree K unit of measure, TC Cold-Junction enabled (in case of TC mode)
and input scale and offset enabled:
The command data is 01010001BIN. = 51 HEX. Then send *W0A51
TABLE 10.2 INPUT CONFIGURATION
INP.CNF and INPUt
3
2
1
0
0
60 Line Frequency 60Hz
1
50 Line Frequency 50Hz
0
SLOW (14/SEC) Reading Rate
1
FAST (100/SEC) Reading Rate
0
UNIPOL Voltage Unipolar; tC
1
BIPOL Voltage Bipolar
0
C Unit of Temperature Degree C
1
F Unit of Temperature Degree F
0
k Unit of Temperature Degree K
1
(Not Used)
tC.CO.JC ENAbLE TC Cold Junction;
3.LINR 3 wire linear RTD;
NORMAL Bridge Limit; Process (Not Used)
tC.CO.JC dISABL TC Cold Junction;
4.LINR 4 wire linear RTD ;
SP1.L2.H SP1 lower overload limit and SP2
upper overload limit; Process - (Not Used)
IN.SC.OF dISABL Input Scale & Offset
IN.SC.OF ENAbLE Input Scale & Offset
RAtIO dISABL Strain Non-Ratiometric;
TC & Process Default
RAtIO ENAbLE Strain Ratiometric