To reset the attenuator contact counter

Turn off the Infiniium oscilloscope.
1
Connect a mouse and a keyboard to the oscilloscope.
2
Turn on the oscilloscope. Insert the service disk when the message "Starting Windows
3
9x..." appears.
appears on the screen.
The oscilloscope will boot into the Windows® operating system.
4
Press Control-Esc on the keyboard to access the Windows® Start menu.
5
Open the Programs group and click the Infiniium Scope icon.
6
When the oscilloscope application comes up, click the mouse icon in the upper right
7
corner of the screen.
Click the Utility pull down menu, then click Self Test. In the Self Test screen click
8
Select Service Extensions.
Click the Self Test pull down menu, scroll down and select Atten Set.
9
Figure 6-33
Self Test
Chapter 6: Replacing Assemblies

To reset the attenuator contact counter

6–33