Lift the probe power and control assembly out and away from the chassis.
5
Figure 6-9
Remove the Probe Power and Control Assembly
To replace the probe power and control assembly, reverse the above procedure.
6
When inserting the assembly, be sure the two tabs on the circuit board engage the two slots in
the sheet metal. Also, be sure to carefully lock in the connector for the mylar flex cable when
reattaching the cable. See "To disconnect and reconnect mylar flex cables" in this chapter.
To remove and replace the probe power and control assembly
Chapter 6: Replacing Assemblies
6–9