Mirus for H2 Classic GrainGage

4.4.4 Chamber Calibration

To create a new moisture curve in Mirus for the H2 Classic GrainGage, do the following:
1. If the grain samples have been in air conditioning or cold storage, acclimate them by spreading them
out on a clean surface in the shop overnight. Do not leave them in the sun to warm up.
2. Test the ambient samples in a benchtop sensor by cycling sub-samples from each larger sample 3–5
times and then averaging the moisture and test weight readings. This is the Known Moisture and
Test Weight.
Note: A test weight from a known source is only needed from one sample.
3. Within the hour, take the samples out to the GrainGage. If you are calibrating outside, make sure the
samples stay in the shade.
H2 Classic GrainGage Calibration
4. Open Mirus.
5. Make sure that the H2 GrainGage plugin is loaded.
6. Select Setup > H2 GrainGage.
7. Scroll down and select Moisture Curves.
8. Tap the new icon
.
54