Reference Manual
00809-0100-4774
7. Set damping: PRIMARY_VALUE_DAMPING.
8. Set up LCD display.
9. Review transmitter configuration.
10. Set switches and software write lock.
Prerequisites
See
Figure 2-13
Before beginning configuration, you may need to verify the device tag or deactivate
hardware and software write protection on the transmitter. To do this, follow the steps
below. Otherwise continue at
1. To verify the device tag:
2. To check the switches (see
Note
Place the control loop in Manual mode before beginning analog input block configuration.
Note
Always check and reconcile function block configuration (with the exception of resource
and transducer blocks) after commissioning the transmitter to the control host. You may
not save function block configuration, including AI blocks, made prior to device
commissioning to the control host to the control host database during the commissioning
process. In addition, the control host may download configuration changes to the
transmitter as part of the commissioning process.
Note
Typically, make changes to the AI block configuration after the transmitter is
commissioned using the control host configuration software. Consult your host system
documentation to see if the AI Block guided configuration method provided in the DD or
DTM should be used after the device has been commissioned.
Note
For DeltaV users, only make final AI block configuration and AI block configuration
changes using the DeltaV Explorer.
Procedure
1. Begin configuring the AI block.
Reference Manual
to graphically view the step by step process for basic device configuration.
Step
a. Navigation: from the Overview screen, select Device Information to verify the
device tag.
Figure
a. The Write Lock switch must be in the unlocked position if the switch has
been enabled in software.
b. To disable the software write lock (devices ship from the factory with the
software write lock disabled):
From the Overview screen, select Device Information and then select the
Security and Simulation tab.
Perform Write Lock Setup to disable software write lock.
1.
1):