Emerson Rosemount 3051CF Series Reference Manual page 32

Pressure transmitter with foundation fieldbus protocol
Hide thumbs Also See for Rosemount 3051CF Series:
Table of Contents

Advertisement

Configuration
July 2020
determine the alarm priorities assigned to each status signal category by your host.
Manual configuration may be required. DeltaV assigns a priority of two or higher. The
priority is based on status signal category. The status signal priority determines the
behavior of both real and simulated alerts.
2.6.4
Plantweb alerts overview
Alerts are generated, mapped, and masked as NE 107 Status Signals. If Plantweb
are required the NE 107 Status Signals are automatically converted to Plantweb alerts for
annunciation and display. Plantweb alerts have three alert parameters. They are in order
from highest to lowest priority:
1. FAILED_ALM
2. MAINT_ALM
3. ADVISE_ALM
The eight alert conditions are factory configured to annunciate as one of the three specific
alert parameters.
Plantweb alert parameter conditions and factory defaults
Emerson ships devices from the factory with all applicable Plantweb
alert conditions reported in each parameter are:
1. FAILED_ALM
2. MAINT_ALM
3. ADVISE_ALM
An ADVISE_ALM indicates a transducer block is not in Auto mode. This may be due to
configuration or maintenance activities. It can also indicate an abnormal process or device
32
a. Incompatible module
b. Sensor failure
c. Electronics failure
A FAILED_ALM indicates a failure within a device that will make the device or some
part of the device non-operational. This implies that the process variable may no
longer be available and the device is in need of immediate repair.
a. Pressure out of limits
b. Sensor temperature out of limits
A MAINT_ALM indicates that the device is experiencing pressure or temperature
conditions that are outside the device operating range. This implies that the
process variable may no longer be accurate. It also implies that if the condition is
ignored the device will eventually fail. The device should be checked to determine
the type of abnormal condition and recommended actions to resolve it.
a. Function check
b. Display update failure
c. Variation change detected
Reference Manual
00809-0100-4774
alerts