CR 35-X D
IGITIZER

Maintenance

Preventive maintenance
Regular preventive maintenance needs to be done once per year or after
12000 cycles (whatever comes first).
This maintenance cannot be done by the user but has to be done by an Agfa
certified field service engineer.
Not performing the regular maintenance by appropriately certified people
can have impact on warranty commitments.
Recurrent safety tests
The digitizer shall be tested according to IEC 62353* in a time interval of at
least 36 months or less if local regulations are different.
*Medical electrical equipment – Recurrent test and test after repair of medical
electrical equipment.
24
Maintenance
4454C EN 2013-05-14