Status field:
Error field:
MESSAGE 1
POWER SUPPLY OUT OF
TOLERANCE
IP JAM
Status field:
MESSAGE 1
ERASURE LAMP
[X], [Y], [Z] DEFECTIVE
SCANNER WARNING
[PPNAME] NOT READY
CORRUPTED MESSAGE IN
QUEUE
UNKNOWN DESTINATION
[PPNAME]
ERROR WHILE LOADING
LANGUAGE FILE
66
Troubleshooting checklist
ERROR
'CODE XXXXX
'
Message 2
1. Check setting of
voltage selector switch
on back panel.
2. Check fuses of the
machine
3. Check supply
voltage.
1. Remove right side
panel
2. Put plate back into
cassette.
3. Close right side panel.
WARNING
Message 2
Press
to complete, IP
is not erased
Possible bad image,
press
Please check and press
Please press
and
check queue
Please press
and
check queue
Default language is
used, please press
Action
Refer to
'Checking the voltage supply'
on page
116
of the Reference manual.
Refer to
'Solving image plate and
cassette jams'
on page
119
Reference manual.
Action
• Press Confirm key.
• Refer to
'Replacing the erasure
lamps'
on page
111
of the
Reference manual.
• Press Confirm key.
• Contact your local service
organization.
• Check image processing station.
• If image processing station is
ready, press Confirm key.
• Press Confirm key.
• Refer to
'Consulting the image
transmission queue ('Queue
management')'
on page
Reference manual.
• Press Confirm key.
• Refer to
'Consulting the image
transmission queue ('Queue
management')'
on page
Reference manual.
• Press Confirm key. English will be
used.
• Restart CR 35-X.