AGFA CR 35-X User Manual page 68

Table of Contents

Advertisement

CR 35-X D
IGITIZER
Status field:
MESSAGE 1
24 X 30 CM CALIBRATION
MISSING
SERVICE MODE
CASSETTE SLOT BLOCKED
IMAGE-QUEUE FULL
UNKNOWN DESTINATION
[PPNAME]
RIGHT SIDE PANEL NOT
CLOSED
UNKNOWN IP-TYPE
EMERGENCY DATA NOT
DEFINED
68
Troubleshooting checklist
LOCKED
Message 2
Press
to accept or
Please wait
Remove cassette, press
Check queue
Press
, remove
cassette and identify
Close right side panel
Press
, remove
cassette, call Service.
Emergency keys
disabled, press
Action
• Press Confirm key to treat
24 x 30 cm image plate without
calibration or press Cancel key to
treat cassettes with other formats.
• Contact your local service
organization.
Wait.
• Remove cassette.
• Remove obstructing objects.
• Press Confirm key.
• Refer to
'Consulting the image
transmission queue ('Queue
management')'
on page
Reference manual.
• Check that the CR 35-X is not off
line (Refer to
'The display'
page 32).
• Press Confirm key.
• Remove cassette.
• Identify cassette.
• Re-insert cassette.
• Check the configuration of the
system.
Close the right side panel.
• Press Confirm key.
• Remove cassette.
• Contact your local service
organization.
• Press Confirm key
• Contact your local service
organization.
66
of the
on
4454C EN 2013-05-14

Advertisement