14. Modified Calibration Tool Lab

Description:
Time Required:
Learning
Objective(s):
Support
Materials
Required:
Tools Required:
Safety
Considerations:
Steps:
Review:
MR5047 SV
This lab will familiarize the student with the modified calibration tool
in SV system, including LV Shim, DQA Tool II, Grafidy 3 and EPI white
pixel test.
2.5 hours
Navigate the service documentation
Practice the modified calibration tools
Service Methods procedures 5364191 on disk, or as loaded onto
system computer.
LV Shim Phantom Assembly
Nesting Plate Assembly
DQA III Phantom
Grafidy 3 Kit
Split Head Coil
1. Follow all LOTO procedures.
incorporated in the documented procedures
2. Take care when moving LV Shim phantom.
1. Use Modified Calibration Tool
Use documented procedure in Direction 5364191:
Installation / Install and Calibration Wizard for Signa Brivo
MR355/Optima MR360 / 4 Procedure / 4.4 Calibrations Guided
Flow
a. Perform a saveInfo to save the current calibration
b. Perform LV Shim
c. Perform DQA Tool II
d. Perform Grafidy 3
e. Perform EPI White Pixel test
f. Restore the saveInfo of the system
Questions for instructor:
______________________________________________________
______________________________________________________
______________________________________________________
They are posted in the lab and
Lab Guide