GE OPTIMA MR360 Service Training page 202

Atc mr instructor xiao xiao
Table of Contents

Advertisement

GE Healthcare
MR5047 SV
k. Record results: ____________________________________
l.
Use command mgd_term to open SCP and AGP terminal
m. Do the TPS Reset and check what will happen
n. Restore the cable
o. Disconnect RUN M1515 from J103 of Mega Switch
p. Perform RF Calib
q. Record results: ____________________________________
r. Use command mgd_term to open SCP and AGP terminal
s. Do the TPS Reset and check what will happen
t. Restore the cable
3. DTX -> RRX DP
Use documented procedure in Direction 5364191:
Troubleshooting/Diagnostics/System Function/RF/Transmitter
Diagnostics
a. Perform DTX -> RRX DP Diagnostics
b. Record results: ____________________________________
c. Disconnect RUN M1513 from J106 of Mega Switch
d. Perform DTX -> RRX DP Diagnostics
e. Record results: ____________________________________
f. Restore the cable
4. Observe Error Message
Refer to the following diagram
a. Record the status of LEDs on the IRF3 board and RRX
b. Disconnect IRF3 J8
c. Note the status of LEDs on the IRF3 board and RRX
d. Record error code: ____________________________________
e. Restore IRF3 J8
f. Disconnect IRF3 J11
g. Note the status of LEDs on the IRF3 board and RRX
Lab Guide
J21
To VRF
Page: 8 of 24