GE Healthcare
11. Fixed-Table Installation & Adjustment Lab
Description:
Time Required:
Learning
Objective(s):
Support Materials
Required:
Tools Required:
Safety
Considerations:
Steps:
MR5047 SV
This lab will familiarize the student with the adjustments needed to
align the table to the gantry and insure all safety/interlock
mechanisms are operational
2 Hour
Navigate the service documentation
Dock Frame Installation
Level table and adjust top height
Check and adjust lock and latch adjustments
Check and adjust cradle emergency latch
Service Methods procedures – 5364191 on disk, or as loaded onto
system computer.
Standard non-magnetic hand tools
Follow all LOTO procedures. They are posted in the lab and
incorporated in the documented procedures
1. Dock Frame Installation
Use documented procedure in Direction 5364191:
Installation / Guided Mechanical Install Flow for Optima MR360
/ Brivo MR355 / Illustration 2: 1st day / Dock Frame Assy
(http://3.28.123.26:8000/mr/cdrom/5364191-
2EN/root/data/Signa_EXCITE/content/1097184.htm)
a. Perform Dock Frame Installation Procedure
2. Top Height Adjustment
Use documented procedure in Direction 5364191:
Adjust/Cals/ Fixed Table / TOP HEIGHT ADJUSTMENT
a. Perform Procedure steps 1 - 7
3. Cradle Side Lock Adjustment
Use documented procedure in Direction 5364191:
Adjust/Cals/ Fixed Table / CRADLE SLIDE LOCK ADJUSTMENT
a. Perform Procedure steps 1 - 8
4. Cradle Guide Latch
Use documented procedure in Direction 5364191:
Adjust/Cals/ Fixed Table / CRADLE GUIDE RAIL LATCH
ADJUSTMENT
a. Perform Procedure 4.1, steps 1 - 4
5. Cradle Emergency Release
Lab Guide
Page: 20 of 24