GE Healthcare
7. Coolant Pumpout Lab
Description:
Time Required:
Learning
Objective(s):
Support Materials
Required:
Tools Required:
Safety
Considerations:
Steps:
Questions:
Review:
MR5047 SV
This lab will familiarize the student with the procedure to pump out
coolant from SRFD3 and XFD in SV system.
0.5 hour
Navigate the service documentation
Understand the importance of coolant pumpout before
components replacement
Practice the correct procedure of coolant pumpout
Service Methods procedures – 5364191 on disk, or as loaded onto
system computer. Refer to Coolant Pumpout procedures.
(http://3.28.123.26:8000/mr/cdrom/5364191-
2EN/root/data/Signa_EXCITE/content/1093487.htm)
Coolant Tank, Water tray, Hand Pump and Standard hand tools
Follow all LOTO procedures as required. They are posted in the lab
as well as incorporated as subset of documented procedure.
After draining, the drain valves must be closed. Check that drain
valves are closed.
Avoid coolant splashing into the electronics.
1. Coolant pumpout
Use documented procedure in Direction 5364191:
Replacement / System Cabinet Replacement / XFA and XFD-PS
Replacement / XFD-PS / 3 Procedure / 3. Drain the coolant
a. Perform the draining operation
1. What is the use of water tray?
__________________________________________________________
2. Why the coolant in the water-cooling components must be drained
when replacement?
__________________________________________________________
3. What action should be executed before restoring the drain hose to
drip pan?
__________________________________________________________
Questions for instructor:
______________________________________________________
______________________________________________________
______________________________________________________
Lab Guide
Page: 14 of 24