Express Coil Lab - GE OPTIMA MR360 Service Training

Atc mr instructor xiao xiao
Table of Contents

Advertisement

GE Healthcare

12. Express Coil Lab

Description:
Time Required:
Learning
Objective(s):
Support Materials
Required:
Tools Required:
Safety
Considerations:
Steps:
Questions:
Review:
MR5047 SV
This lab will familiarize the student with the setup and image quality
assurance tests for the Express coil.
1 Hour
Navigate the service documentation
Position the phantoms and coils for testing
Perform the Multi-Coil Quality assurance Tool
Service Methods procedures 5364191 on disk, or as loaded onto
system computer. Refer to Express coil SNR test .
(http://3.28.123.26:8000/mr/cdrom/5364191-
2EN/root/data/Signa_EXCITE/content/1095326.htm)
Unified Phantom
Phantom positioners, pad, and strap
Follow all LOTO procedures. They are posted in the lab and
incorporated in the documented procedures
1. Express Coil MCQA Test
Use documented procedure in Direction 5364191:
Adjust/Cals / Surface and Head Coils / 1.5T Coil Vendor Manuals
/ Table 11: Express Coil / SNR Test
a. Perform procedure 3.1 step 1 –3 to setup HNA coil and position
phantom
b. Perform step 4 to run MCQA tool to test mode HNA+PA
c. Perform procedure 3.1 step 1 –4 to setup AA coil and position
phantom
d. Perform step 5 to run MCQA tool to test mode AA+PA
e. Compare and record test results
1. What coil segment would be indicated if there was a failure test 2,
Sig_Img 4 in mode HNA+PA ?
__________________________________________________
2. What coil segment would be indicated if there was a failure test 2,
Sig_Img 4 in mode AA+PA ?
__________________________________________________
Questions for instructor:
______________________________________________________
______________________________________________________
______________________________________________________
Lab Guide
Page: 22 of 24