Icn Diagnostic Lab - GE OPTIMA MR360 Service Training

Atc mr instructor xiao xiao
Table of Contents

Advertisement

GE Healthcare

5. ICN Diagnostic Lab

Description:
Time Required:
Learning
Objective(s):
Support
Materials
Required:
Tools Required:
Safety
Considerations:
Steps:
MR5047 SV
This lab will familiarize the student with the function of three
ethernet cables connected to ICN.
1.5 Hour
Understand the function of NET MGT, Port2 and Port 3 of ICN
Service Methods procedures – 5364191 on disk, or as loaded onto
system computer. Refer to ICN Reconfiguration procedure.
Standard hand tools
Follow all LOTO procedures as required. They are posted in the lab
as well as incorporated as subset of documented procedure.
1. Understanding Ethernet cables function
a. Open CShell window and input more /etc/hosts
b. Record the IP address of icn1 vre and icn1_bmc
c. Start a new Exam to scan
d. Disconnect the cable connected to NET MGT port of the ICN
e. Open a C-shell window and ping icn1 and icn1-bmc.
f. Start to scan
g. Disconnect the cable connected to Port2 of the ICN
h. Ping icn1 and icn1-bmc
Lab Guide
Page: 11 of 24

Hide quick links: