• Fixed table has no Dock and Undock function like detachable table. Up and down of fixed
table are implemented by actuator, which is driven by electrical motor. The table can go
up and down very smoothly with very low noise compared to the detachable table. And
there is no other mechanism to drive fixed table up and down.
• The control box for fixed table is newly designed.
• The longitudinal movement of the cradle is driven by longitudinal motor located in the
rear pedestal. It is same as HDe.
• The cover of the cradle is reinforced to protect the embedded Express PA coil from
damage. The cradle with the Express PA coil moves the patient into and out of the
magnet bore.