GE OPTIMA MR360 Service Training page 109

Atc mr instructor xiao xiao
Table of Contents

Advertisement

2. Output control signal:
a. E-Off Contactor Trip Control
b. Night Mode Contactor Trip Control
c. Main Breaker Trip Control: In HDe system, when leak sensor 1 detects any leakage,
cabinet monitor will send control signal to PDU to trip the Main Breaker. The whole
system will lost power input. But in SV system, leak sensor 1 is removed. So this signal
is not used in SV any more although the control circuit is still there.
107

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Brivo mr355

Table of Contents