GE OPTIMA MR360 Service Training page 132

Atc mr instructor xiao xiao
Table of Contents

Advertisement

Performance:
Water cooling gradient system
Slew Rate: 100 T/m/s
G-Max: 33 mT/m
Input : 208VAC, 1A ; 200VDC&700VDC
Output: 300A for 100ms, 1400VDc
Cal/Function Check:
DC offset, DQA cal
Diagnostic:
Hammer, Frame/Clk test, ECC Verification, Hysteresis Static Test is obsolete.
FRU: XFA, XFD-PS, Control Boards
Install/Replacement:
1) Quick disconnecting design to facilitate the replacement.
2) Need to remove the leak sensor first to replace control board.
3) Super capacitor design of XFD-PS require ~30 minutes to discharge.
4) 87 Kg weight of XFD-PS. Need hoist tool