GE OPTIMA MR360 Service Training page 61

Atc mr instructor xiao xiao
Table of Contents

Advertisement

• For the purpose of market sales, Brivo 355 has four kinds of configuration: Main, Main +
Knee, Main + Shoulder and Premium.
• In Brivo Main configuration, only 8 coils listed in the left table can be used.
• In Brivo Main + Knee configuration, not only the initial 8 coils but also Quad Extremity coil
can be used.
• In Brivo Main + Shoulder configuration, the 8 initial coils and 3 channel shoulder coil can
be used.
• In Brivo Premium configuration, 10 coils can be used, including the Quad extremity, 3
channel shoulder coil and the 8 coils in Main configuration.
• So, only the coils included in the configuration can be used. For example, 3Ch Shoulder
coil can't be used in Brivo Main and Brivo Main + Knee configurations. Be sure to confirm
what kind of configuration is in your site before using any coils, otherwise the coil may be
not identified.
• There is no this limitation in Optima MR360. But MR360 has to follow another limitation
according to different table type.(Refer to the next page)