There are some steps during the installation of the MR system that must be done but for
which the system has no "machine" data. For these items the system will rely on the user to
check off the items as they are complete and click [Submit].
For all calibrations that have machine data, the tool can be launched the same way it was
launched in previous versions of MR software – by clicking [Click here to start this tool]. The
tool will launch in a separate window. After completion of the tool, the user must click
[Check Status].
After clicking [Submit] or [Check Status], the ICW will check the data files. If it finds a passed
calibration or a completed checklist, then it will allow the user to advance to the next
calibrations in the flow. If it detects a failure or an incomplete task, it will turn the name of
the tool RED in the left window of the ICW and no new calibrations will become available
until that step has been completed successfully.