Invacare 2G Tarsys Service Manual page 37

Table of Contents

Advertisement

TABLE 10 - 2GT and 2GTR Systems WITHOUT Center Mount Footrests AND
SEAT
DEPTH
20-INCH
*NOTE: Linear guides are exploded for clarity. DO NOT disconnect the upper
21-INCH
and lower portions of the linear guides. The linear guide cannot be reassembled.
22-INCH
Tilt Actuator
Rear Pivot Assembly
Linear Guide*
Socket Screw
Seat Bracket
Lower Interface
Bracket
NOTE: Torque the hardware as listed below:
• Hex screws or socket screws (securing the rear pivot): 75 in-lbs ± 20%.
• Hex screws (securing the front pivot assembly and front of seat bracket) and shoulder
screw (securing the tilt actuator): 13 ft-lbs ± 20%.
Part No 1114842
Ventilator-Ready Battery Tray
MOUNTING POSITION
Shim
Washer
Locknuts
Locknut
Hex Screw
SECTION 5—INTERFACE HARDWARE
Locknuts
(STEP 9)
37
2G Tarsys®Powered Seating System
Shoulder Bolt
and Locknut
Front Pivot
Assembly
Hex Screws