38
D750 master board for Mammomat Novation
Fig. 13: D750
•
Turn the system OFF.
•
Remove the right stand cover.
•
Remove board D750 from the stand.
•
IC 45, IC 53, IC 62, IC 52 have to be moved from the old to the new
board(1/Fig. 13 / p.
•
Check the position of switch S5 and the bridge on ...X754 1-2
•
Install the new board.
•
Install the right stand cover.
•
Turn the system ON.
•
No backup/restore or adjustment needed.
MAMMOMAT Novation DR
38).
SPB7-250.841.01.04.02
05.05
Replacing Stand Boards
(2/Fig. 13 / p.
Page 38 of 64
CS SD 24
0
38).
Siemens AG
Medical Solutions