Site Preparations

Invenia ABUS 2.0 environmental requirements

If the Invenia ABUS 2.0 is very cold or hot
CAUTION
Cooling
NOTE:
Lighting
2-2

General requirements

When unpacking the Invenia ABUS 2.0, allow the temperature
of the Invenia ABUS 2.0 to stabilize before powering up. The
following table describes guidelines for reaching operational
temperatures from storage or transport temperatures.
If the Invenia ABUS 2.0 is very cold or hot, do not turn on its
power until it has had a chance to acclimate to its operating
environment.
Table 2-1:
Invenia ABUS 2.0 acclimation time
The cooling requirement for a typical console Ultrasound system
with monitor and on board peripherals, is up to 3800 BTU/h.
This figure does not include cooling needed for lights, people, or
other equipment in the room.
Each person in the room places an additional 300 BTU/h
demand on the cooling system.
Bright light is needed for Invenia ABUS 2.0 setup, updates, and
repairs. However, operator and patient comfort may be
optimized if the room light is subdued and indirect. Therefore a
combination lighting system (dim/bright) is recommended. Keep
in mind that lighting controls and dimmers can be a source of
Electromagnetic Interference (EMI) which could degrade image
quality. These controls should be selected to minimize possible
interference.
Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4