Common Error Codes For The Invenia Abus 2.0 - GE Invenia ABUS 2.0 Technical Publication

Table of Contents

Advertisement

Common Error Codes for the Invenia ABUS 2.0

Error Code (total 5 digits)
First 1 digit
1 - ABUSApp
related
2 - ScannerApp
related
3 - GEUS related
NOTE:
Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
The Invenia ABUS 2.0 system error messages have been
improved to provide an easier troubleshooting experience for
field engineers and remote technical support team members.
The Invenia 2.0 system displays an error/information message
along with an error code. The error code displayed uses the
same error code rule (5 digit number) implemented with the
Invenia R1 system:
Second 1 digit
1 - Acquisition related
2 - Scanhead related
3 - Renderer related
4 - System status related
1 - Initialization related
2 - USB related
1 - Failure Type
2 - Info Type
3 - Service Type
The software team ensured that the new GEUS (Ultrasound
Box) error codes does not overlap with old Verasonics (VDAS/
Acquisition) codes. Thus, the error codes related to GEUS and
VDAS are unique to each system design.
Invenia ABUS 2.0 Error Log
Third 3 digits
Acquisition error code
001 - Scanhead id does not match.
001 - Image renderer initialization
failed.
ScannerApp initialization related error
code
ScannerApp USB related error code
after initialization
GEUS fatal error information category
(requires system shutdown)
GEUS information error information
category (warning/information)
GEUS service error information
category (pass to service app)