Service Configuration Menu - GE Invenia ABUS 2.0 Technical Publication

Table of Contents

Advertisement

System Setup

Service Configuration Menu

Table 3-7:
Service Task
Delete or anonymize patient
information
Keep 16-bit data
Export Anonymization
GPU
Export and Import Configuration
DICOM Connection
3-68
From the Configuration Menu (Gears Icon), select Service.
The Service Configuration Menu appears.
Figure 3-29. Service Configuration Menu
Service Configuration Menu Functionality
Clean System: Completely removes all patient data out of the image
data folder. Deletes any study on the Invenia ABUS 2.0.
ALWAYS set this to OFF.
On/Off -- When you export patient data to a USB drive, patient
demographic information is removed.
• Speed -- Speed of the GPU
• Temperature -- Temperature of GPU and Motherboard
• Export to USB -- Exports system configuration settings to USB.
• Import from USB -- Imports system configuration settings to USB.
NOTE: Reboot the system after importing the configuration.
• Worklist -- DICOM Worklist Verify
• MPPS -- DICOM MPPS Verify
• Workstation DICOM Workstation Verify
Invenia ABUS 2.0 – System Setup and Basic Service Manual