Electrical requirements

General requirements
NOTE:
CAUTION
Site power outlets
Unit power plug
Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
The Invenia ABUS 2.0 requires a dedicated power and ground
for the proper operation of its Ultrasound equipment. A separate
power outlet with a minimum 10 amp circuit breaker for
100-240 VAC.
Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Invenia ABUS 2.0 is only a conduit.
Power outage may occur. The Invenia ABUS 2.0 requires a
dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you do not
have any other equipment operating on the same circuit.
A dedicated AC power outlet must be within reach of the Invenia
ABUS 2.0 without extension cords. Other outlets adequate for
the external peripherals, medical and test equipment needed to
support this Invenia ABUS 2.0 must also be present within 1 m
(3.2 ft.) of the Invenia ABUS 2.0. Electrical installation must
meet all current local, state, and national electrical codes.
Refer to Chapter 9 for Power Plug part numbers. If the Invenia
ABUS 2.0 arrives without a power plug, or with the wrong plug,
you must contact your GE dealer or the installation engineer
must supply what is locally required.