Functional Checks

Overview

Performance Functional Checks

Power On/Boot Up Functional Check
Power Shut Down Functional Check
Transducer Functional Check
4-12
This section describes the functional checks used with the
Invenia ABUS 2.0. Functional checks are used to verify that the
product works as specified after performing maintenance
procedures. Functional checks may also be used during
troubleshooting.
Confirm that the system:
Powers on when you press the On/Off/Standby control.
That the transducer moves across the Scan Head
Assembly.
That the software runs through its initialization.
And that each user type (User or Admin) can successfully
login to the system.
Confirm that the system exits the application and powers down.
Ensure that the transducer:
Upon Power Up moves across the Scan Head Assembly.
That the lights on the Scan Head Assembly light up.
That the Transducer images as a scan is being performed.
That the Transducer generates a 3D Volume.
Invenia ABUS 2.0 – System Setup and Basic Service Manual
Functional checks
4700-0043-00 Rev. 4