Ball Joint Lubrication - GE Invenia ABUS 2.0 Technical Publication

Table of Contents

Advertisement

Service Adjustments and lubrication

Ball Joint Lubrication

1. Apply a thin film of lubricant to lower half of the nickel-plated ball joint while tipping Scan
Head in all four directions. Clean off any excess lubricant from the painted surfaces.
2. Remove the:
• Scan Head Membrane, if in place•
• Scan Head Cassette from the Scan Head
• Clear Cast Cover
3. Rest the Scan Head Assembly on a flat, safe surface. The Scan Head Assembly must be supported for
this procedure.
6-4
Perform this procedure only if the ball joint is squeaking.
The following lubricant is required to grease the nickel-plated
ball joint: DuPont™ Krytox® 227 Synthetic Performance
Lubricant (P/N 2301-0478-00)
Table 6-2:
Ball Joint Lubrication Points
Steps
Invenia ABUS 2.0 – System Setup and Basic Service Manual
Illustration
4700-0043-00 Rev. 4