SP4T Raptor Holder Assembly Removal

Tools Required
TORX screwdriver, T10, T15 and T20.
Box wrench 5.5mm
Torque wrench, 5/8 in.
Procedure
Refer to Figure 5-9 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 178.
Step 2. Detach the front chassis as described in "Front Chassis Detachment" on page 179.
Step 3. Remove the front frame as described in "Front Frame Removal" on page 180.
Step 4. Remove the assembly board as described in "Main Board Removal" on page 184.
Step 5. Remove the SP4T switch assembly as described in "SP4T Switch Assembly Removal" on
page 185.
Step 6. Remove the two TORX T10 screw (item 1) on the SP4T raptor holder assembly and
remove the cables.
Step 7. Remove the four TORX T10 screw (item 2) holding the SP4T raptor holder assembly and
lift the raptor holder assembly.
Step 8. Remove the semirigid cable (item 3) on the SPDT switch assembly.
Step 9. Remove the six TORX T10 screw (item 4) holding the SP4T raptor holder assembly and
lift the raptor holder assembly.
Chapter 5
Replacement Procedure
Replacement Procedure for E5092A-020
187