Agilent Technologies ENA Series Service Manual page 61

Configurable multiport test set
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

From Port -
Frequency
To Port
Range [Hz]
(Selected Port)
Port 1 - 1D
50M to 500M
(1B)
500M to 1G
1G to 2G
2G to 6G
6G to 10G
10G to 18G
18G to 20G
Port 1 - 1D
50M to 500M
(1C)
500M to 1G
1G to 2G
2G to 6G
6G to 10G
10G to 18G
18G to 20G
From Selected Port To Unselected Port
From Port -
Frequency
To Port
Range [Hz]
(Selected Port)
1A - 1B (1A)
50M to 500M
500M to 1G
1G to 2G
2G to 6G
6G to 10G
10G to 18G
18G to 20G
Chapter 2
E5092A Performance Test Record
Test
Test Result [dB]
Limit
[dB]
< -65.0
< -80.0
< -85.0
< -90.0
< -85.0
< -75.0
< -65.0
< -65.0
< -80.0
< -85.0
< -90.0
< -85.0
< -75.0
< -65.0
Test
Test Result [dB]
Limit
[dB]
< -65.0
< -80.0
< -85.0
< -90.0
< -85.0
< -75.0
< -65.0
Performance Tests
Measurement
Uncertainty
[dB]
± 0.35
± 0.48
± 0.60
± 0.76
± 0.66
± 0.54
± 0.43
± 0.35
± 0.48
± 0.60
± 0.76
± 0.66
± 0.54
± 0.43
Measurement
Uncertainty
[dB]
± 0.35
± 0.49
± 0.62
± 0.71
± 0.71
± 0.44
± 0.36
55

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5092a

Table of Contents