Agilent Technologies ENA Series Service Manual page 85

Configurable multiport test set
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Port
Frequency
(Selected
Range [Hz]
Port)
3D (3B)
50M to 3G
3G to 10G
10G to 16G
16G to 18G
18G to 20G
3A (3C)
50M to 3G
3G to 10G
10G to 16G
16G to 18G
18G to 20G
3B (3C)
50M to 3G
3G to 10G
10G to 16G
16G to 18G
18G to 20G
3D (3C)
50M to 3G
3G to 10G
10G to 16G
16G to 18G
18G to 20G
3A (3D)
50M to 3G
3G to 10G
10G to 16G
16G to 18G
18G to 20G
Chapter 2
E5092A Performance Test Record
Test Limit
Test Result [dB]
[dB]
< -17.0
< -11.0
< -8.0
< -6.0
< -4.0
< -17.0
< -11.0
< -8.0
< -6.0
< -4.0
< -17.0
< -11.0
< -8.0
< -6.0
< -4.0
< -17.0
< -11.0
< -8.0
< -6.0
< -4.0
< -17.0
< -11.0
< -8.0
< -6.0
< -4.0
Performance Tests
Measurement
Uncertainty
[dB]
± 0.43
± 0.37
± 0.28
± 0.28
± 0.35
± 0.43
± 0.37
± 0.28
± 0.28
± 0.35
± 0.43
± 0.37
± 0.28
± 0.28
± 0.35
± 0.43
± 0.37
± 0.28
± 0.28
± 0.35
± 0.43
± 0.37
± 0.28
± 0.28
± 0.35
79