Main Board Removal - Agilent Technologies ENA Series Service Manual

Configurable multiport test set
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Replacement Procedure
Replacement Procedure for E5092A-020

Main Board Removal

Tools Required
Procedure
Refer to Figure 5-7 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 178.
Step 2. Detach the front chassis as described in "Front Chassis Detachment" on page 179.
Step 3. Remove the front frame as described in "Front Frame Removal" on page 180.
Step 4. Remove the three TORX T10 screws (item 1) from the assembly.
Step 5. Remove the four hex nut (item 2) from the assembly.
Step 6. Lift the main board from the assembly.
Figure 5-7
Main Board Removal
184
Box wrench 5.5mm
TORX screwdriver, T10, T15 and T20.
Chapter 5

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5092a

Table of Contents