From Port
Frequency
- To Port
Range [Hz]
Port 4 - 4B
50M to 100M
100M to 2G
2G to 3G
3G to 4G
4G to 6G
6G to 8G
8G to 10G
10G to 14G
14G to 18G
18G to 20G
Port 4 - 4C
50M to 100M
100M to 2G
2G to 3G
3G to 4G
4G to 6G
6G to 8G
8G to 10G
10G to 14G
14G to 18G
18G to 20G
Chapter 2
E5092A Performance Test Record
Test
Test Result [dB]
Limit
[dB]
> -4.0
> -3.5
> -4.5
> -5.5
> -6.0
> -7.5
> -8.5
> -9.5
> -10.5
> -12.0
> -4.0
> -3.5
> -4.5
> -5.5
> -6.0
> -7.5
> -8.5
> -9.5
> -10.5
> -12.0
Performance Tests
Measurement
Uncertainty
[dB]
± 0.27
± 0.24
± 0.18
± 0.19
± 0.11
± 0.09
± 0.24
± 0.29
± 0.29
± 0.63
± 0.27
± 0.24
± 0.18
± 0.19
± 0.11
± 0.09
± 0.24
± 0.29
± 0.29
± 0.63