suspect.
If the voltage of ADC2 is not within the limit, the A1 board or the heater is suspect.

Check the voltage output of the Control Lines

Step 1. Turn on the E5092A.
Step 2. Measure the DC voltage between the pin 12 and 15 on the control lines using a voltmeter
with a small probe. Figure 3-5 shows the pin assignment of the control lines.
Figure 3-5
Pin assignment of the control lines
Step 3. Rotate the voltage adjustment trimmer on the front panel. Check that the trimmer can set
the voltage from +2V to +5V.
If the output voltage can not be varied from +2V to +5V though the A1 board works
correctly, the A3 DUT control board is suspect.
If the output voltage is good, the A3 board is verified.

Check control lines setting

The E5092A can control the output from the control line (Figure 3-5) of the E5092A and
control the DUT (for example, switching the frequency band of the front end module).
Step 1. Short between the pin 12 - 13 and the pin 14 - 15.
Step 2. Press
Step 3. Press
the real panel of the E5092A.
Step 4. Press
line.
Chapter 3
-
Multiport Test Set Setup
and then select the ID of the E5092A The ID is set with the bit switch on
Test Set 1
Control Lines
from group A until group D to the setting menu of the DUT control
E5092A Setting Troubleshooting
to display the Multiport Test Set Setup menu.
Troubleshooting