Agilent Technologies ENA Series Service Manual page 127

Configurable multiport test set
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

From Selected Port To Unselected Port
Frequency
From Port
Range [Hz]
- To Port
(Selected
Port)
7A - 7B
50M to 500M
(7A)
500M to 1G
1G to 2G
2G to 6G
6G to 10G
10G to 18G
18G to 20G
7B - 7A
50M to 500M
(7B)
500M to 1G
1G to 2G
2G to 6G
6G to 10G
10G to 18G
18G to 20G
Chapter 2
E5092A Performance Test Record
Test Limit
Test Result [dB]
[dB]
< -65.0
< -80.0
< -85.0
< -90.0
< -85.0
< -75.0
< -65.0
< -65.0
< -80.0
< -85.0
< -90.0
< -85.0
< -75.0
< -65.0
Performance Tests
Measurement
Uncertainty
[dB]
± 0.35
± 0.49
± 0.62
± 0.71
± 0.71
± 0.44
± 0.36
± 0.35
± 0.49
± 0.62
± 0.71
± 0.71
± 0.44
± 0.36
121