Agilent Technologies ENA Series Service Manual page 57

Configurable multiport test set
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

From
Frequency
Port - To
Range [Hz]
Port
Port 1 -
50M to 100M
1B
100M to 2G
2G to 3G
3G to 4G
4G to 6G
6G to 8G
8G to 10G
10G to 14G
14G to 18G
18G to 20G
Port 1 -
50M to 100M
1C
100M to 2G
2G to 3G
3G to 4G
4G to 6G
6G to 8G
8G to 10G
10G to 14G
14G to 18G
18G to 20G
Chapter 2
E5092A Performance Test Record
Test Limit
Test Result [dB]
[dB]
> -4.0
> -3.5
> -4.5
> -5.5
> -6.0
> -7.5
> -8.5
> -9.5
> -10.5
> -12.0
> -4.0
> -3.5
> -4.5
> -5.5
> -6.0
> -7.5
> -8.5
> -9.5
> -10.5
> -12.0
Performance Tests
Measurement
Uncertainty
[dB]
± 0.27
± 0.24
± 0.18
± 0.19
± 0.11
± 0.09
± 0.24
± 0.29
± 0.29
± 0.63
± 0.27
± 0.24
± 0.18
± 0.19
± 0.11
± 0.09
± 0.24
± 0.29
± 0.29
± 0.63
51

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5092a

Table of Contents