Troubleshooting
E5092A Setting Troubleshooting
Step 5. Use the corresponding softkey to set the control line of the E5092A.
Softkey
,
,
Line 1
Line 2
Line 3
,
Line 7
Line 8
Step 6. Measure the DC voltages of the control lines using a voltmeter with a small probe. If
is set for each control line, the voltage of the control line is the same value as the output
voltage of the pin 12. If
same value as the output voltage of the pin 15.
•
If any of the voltages of the control lines are not the same voltages as the pin 12 or pin
15, the A3 board is suspect.
•
If all the voltages of the control lines are good, the A3 board is verified.
144
Function
,
,
,
,
Set
Line 4
Line 5
Line 6
is set for each control line, the voltage of the control line is the
Low
/
of each line of the control line.
High
Low
High
Chapter 3