Table 6-1 Post Repair Procedures lists the required procedures that must be performed
after the replacement of an assembly. These are the recommended minimum procedures to
ensure that the replacement is successfully completed.
Table 6-1
Post-Repair Procedures
Replaced
Assembly or Part
Switch Control USB PCA
(E5092-66601)
Switch Indicator LED PCA
(E5092-66602)
DUT Control PCA
(E5092-66603)
194
Required
Adjustments
Correction Constants
(CC)
No adjustment needed
No adjustment needed
No adjustment needed
Verification
Perform the following performance tests.
- Write SN
- DC Source Output Voltage Accuracy Test
- DC Source Maximum Current Test
- Control Line Maximum Current Test
- LED Check
- DCBUS Check
Perform the following performance tests.
- Write SN
- DC Source Output Voltage Accuracy Test
- DC Source Maximum Current Test
- Control Line Maximum Current Test
- LED Check
- DCBUS Check
Perform the following performance tests.
- Write SN
- DC Source Output Voltage Accuracy Test
- DC Source Maximum Current Test
- Control Line Maximum Current Test
- LED Check
- DCBUS Check