Figure 3-6 shows the block diagram of the control line.
Figure 3-6
Circuit diagram of the control line

Specifications

Item
Connector shape
Voltage
range
Maximum current
Impedance
Range of variable voltage

Setting the voltage of variable voltage output

The output voltage of the pin 12 can be varied from +2 V to +5 V. To set the vlotage,
perform the following procedure.
Step 1. Turn on the E5091A.
Step 2. Measure the voltage between the pin 12 and 15 by a multimeter.
Step 3. Rotate the voltage adjustment trimmer on the front panel so that a multimeter show the
voltage to be set.
3
Specifications
15-pin femal D-Sub
Positive Input
0 V to +5 V
−5 V to 0 V
Negative Input
50 mA in total of each line
< 10 Ω
+2 V to +5 V
Setting the Instrument
Setting control line