Safety Precautions; Earthquake - Agilent Technologies 4072A Advanced Pre-Installation Manual

Parametric
Hide thumbs Also See for 4072A Advanced:
Table of Contents

Advertisement

Site Preparation

Safety Precautions

Safety Precautions
To maintain safe operating conditions for the 4072/4073 tester, observe the following earthquake, fire, electrical,
and chemical safety precautions.

Earthquake

To prevent the system cabinet from tipping over in an earthquake, attach the system cabinet to the floor as shown in
Figure 2-13 (a). Use suitable metal fittings to attach the system cabinet, taking into account the system cabinet
screw holes and floor conditions.
For example, Figure 2-13 (b) shows the dimensions for suitable metal fittings. These metal fittings and screws are
not furnished with the system. Please prepare these parts. The metal fittings are appropriate for the Uniform
Building Code (UBC) Zone 4.
The system cabinet's center of gravity is the center of the system cabinet, as shown in Figure 2-14. The center of
gravity may be changed when reconfiguring any instruments in the system cabinet.
Figure 2-13
Fixing system cabinet
(a)Attaching System Cabinet
Figure 2-14
System cabinet's center of gravity
32
5(Min)
127
(80)
SCREW (Attach to floor)
SCREW (M8x1.25)
(b)Dimentions of Metal Fitting
Agilent 4072/4073 Preinstallation Guide, Edition 4
486.7
127
182.7
25
4- Æ 8.4 (for M8 Screw)
Attach to floor