Connecting Site Power Line

WARNING
Main power cable connection must be performed by qualified personnel at the customer site.
Incorrect installation will result in serious injury to operating personnel and damage to the
equipment.
NOTE
See "Main power cable" for main power cable specification.
When to connect the power cable to the power distribution unit (PDU), ensure that it is
connected to the 30 A service.
Use the following procedure to connect the power distribution unit (PDU) to the site power line (source).
1. Set the switchboard breaker to OFF.
disconnect
device
emergency
breaker
3. Set the disconnect device and emergency breaker
to OFF.
Appendix A
Agilent 4072/4073 Preinstallation Guide, Edition 4
Confirming the System Power Safety
Connecting Site Power Line
2. Open the system cabinet rear door with the key
provided.
4. Remove the PDU side cover.