Agilent Technologies 4072A Advanced Pre-Installation Manual page 67

Parametric
Hide thumbs Also See for 4072A Advanced:
Table of Contents

Advertisement

Figure 4-3
Testhead cover removing location
Air Shower Room
(sticky sheet)
You must confirm
the temperature and the waiting time.
Table 4-2
Warming time to ensure no condensation occurs in testhead
Atmospheric
temperature at entrance
− 20 ° C to 0 ° C
0 ° C to 10 ° C
10 ° C to 20 ° C
Greater than 20 ° C
Chapter 4
Agilent 4072/4073 Preinstallation Guide, Edition 4
Clean Room
Waiting time
under 70 % humidity in buffer room
10 hours
8 hours
5 hours
1 hour
Moving Cabinet, Testhead and Accessories
Buffer Room
(sticky sheet)
Remove all
vinyl cover
Remove the carton and
outer vinyl cover
Waiting time
under 50 % humidity in buffer room
6 hours
4 hours
1 hour
0 hour
Receiving
67