Safety Precautions

To maintain safe operating conditions for the 4072A/4073A tester, observe the following earthquake, fire,
electrical, and chemical safety precautions.

Earthquake

To prevent the system cabinet from tipping over in an earthquake, fix the system cabinet to the floor as shown in
figure 2-5. (a). Use suitable metal fittings to fix the system cabinet, taking into account the system cabinet screw
holes and floor conditions.
For example, figure 2-5. (b) shows the dimensions for suitable metal fittings. These metal fittings and screws are
not furnished with the system. Please prepare these parts. The metal fittings are appropriate for the Uniform
Building Code (UBC) Zone 4.
The system cabinet's center of gravity is the center of the system cabinet, as shown in figure 2-6.. The center of
gravity may be changed when reconfigurating any instruments in the system cabinet.
Figure 2-5.
Fixing System Cabinet
(a)Fixing System Cabinet
Figure 2-6.
System Cabinet's Center of Gravity
The Center
of Gravity
W = 250 kg
(with 4284A, 3458A,
745/132 L, 81110A x 4,
8114A or 81110A)
Chapter 2
Agilent 4072A/4073A Pre-Installation Guide, Edition 3
 /KP

SCREW (FIX to FLOOR)
SCREW (M8x1.25)
600
268.7
39.2
521.5
486.7
127
182.7
4- Æ8.4 (for M8 Screw)
Fix to Floor
(b)Metal Fitting Dimension
364.0
695.6
143.8
Site Preparation
Safety Precautions
127
25
The Center
of Gravity
23